Detection of hyperoxaemia in neonates: data from three new pulse oximeters
1 Department of Neonatology and Pediatric Pulmonology, Hannover Medical School, Hannover, Germany
2 Department of Neonatology, University Hospital, Tübingen, Germany
Correspondence to:
Correspondence to:
Professor Poets, Department of Neonatology, University of Tübingen, Calwerstr 7, D-72070 Tübingen, Germany;
Christian-F.Poets{at}med.uni-tuebingen.de
Aim: To determine the sensitivity and specificity of three newly developed pulse oximeters in the detection of hyperoxaemia, defined as an arterial partial pressure of oxygen (PaO2) of > 80 mm Hg.
Methods: SpO2 readings from three oximeters (Agilent Viridia (AgV), Masimo SET (MaS), Nellcor Oxismart (NeO)) were documented in 56 infants (median gestational age at birth 35.5 weeks, range 2441) whenever an arterial blood gas was taken for clinical purposes. Blood samples were analysed within one minute in a Radiometer ABL 505 blood gas analyser and OSM3 co-oximeter.
Results: Between 280 and 291 blood gases were analysed for each instrument; 105112 showed a PaO2 > 80 mm Hg. At an upper alarm limit of 95%, the three instruments detected hyperoxaemia with 9395% sensitivity. Specificity at this alarm level ranged from 26 to 45%. The mean (SD) difference between arterial oxygen saturation and SpO2 (bias) was -0.25 (2.5)% for AgV, -0.06 (2.5)% for MaS, and -0.91 (2.6)% for NeO (p < 0.01, NeO v AgV and MaS).
Conclusion: These instruments detected hyperoxaemia with sufficient sensitivity at an upper alarm limit of 95%, but showed differences in their specificity, which was probably related to differences in measurement bias.
Keywords: pulse oximetry; oxygenation; alarm limits
Abbreviations: AgV, Agilent Viridia; MaS, Masimo SET; NeO, Nellcor Oxismart; PaO2, arterial partial pressure of oxygen; SaO2, arterial oxygen saturation; SpO2; pulse oximeter saturation
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
-
Quine, D, Stenson, B J
(2009). Arterial oxygen tension (Pao2) values in infants <29 weeks of gestation at currently targeted saturations. Arch. Dis. Child. Fetal Neonatal Ed.
94: F51-F53
[Abstract] [Full Text] -
Quine, D, Stenson, B J
(2008). Does the monitoring method influence stability of oxygenation in preterm infants? A randomised crossover study of saturation versus transcutaneous monitoring. Arch. Dis. Child. Fetal Neonatal Ed.
93: F347-F350
[Abstract] [Full Text] -
Castillo, A., Sola, A., Baquero, H., Neira, F., Alvis, R., Deulofeut, R., Critz, A.
(2008). Pulse Oxygen Saturation Levels and Arterial Oxygen Tension Values in Newborns Receiving Oxygen Therapy in the Neonatal Intensive Care Unit: Is 85% to 93% an Acceptable Range?. Pediatrics
121: 882-889
[Abstract] [Full Text] -
Hagadorn, J. I., Furey, A. M., Nghiem, T.-H., Schmid, C. H., Phelps, D. L., Pillers, D.-A. M., Cole, C. H., and the AVIOx Study Group,
(2006). Achieved Versus Intended Pulse Oximeter Saturation in Infants Born Less Than 28 Weeks' Gestation: The AVIOx Study. Pediatrics
118: 1574-1582
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.