Motion resistant pulse oximetry in neonates

R Sahni, A Gupta, K Ohira-Kist and T S Rosen

College of Physicians and Surgeons, Columbia University, New York, NY, USA

Correspondence to:
Correspondence to:
Dr Sahni
Department of Pediatrics, College of Physicians and Surgeons, Columbia University, 630 W 168th Street, New York, NY 10032, USA; rs62{at}columbia.edu

Background: Pulse oximetry is widely used in neonates. However, its reliability is often affected by motion artefact. Clinicians confronted with questionable oxygen saturation (SpO2) values often estimate the reliability by correlating heart rate (HR) obtained with the oximeter with that obtained by electrocardiogram.

Objective: To compare the effects of motion on SpO2 and HR measurements made with Masimo signal extraction technology and those made with a Nellcor N-200.

Design: Continuous pulse oximetry and HR monitoring were performed in 15 healthy, term infants (mean (SD) birth weight 3408 (458) g) undergoing circumcision, using Masimo and Nellcor pulse oximeters and a standard HR monitor (Hewlett-Packard). Simultaneous minute by minute behavioural activity codes were also assigned. Baseline data were collected for 10 minutes when the infant was quietly asleep and then continued during and after circumcision for a total duration of one hour. The oximeter HR and SpO2 values were compared and related to HR values obtained by ECG during all three periods. The effect of behavioural activity on SpO2 and HR was also evaluated.

Results: When compared with results obtained with the Nellcor, the mean SpO2 and HR were higher and the incidence of artefact lower with the Masimo during all three periods. Masimo HR more accurately predicted HR obtained with a standard monitor, with lower residual error. SpO2 and HR values obtained with the Nellcor were lower and more variable during all behavioural states, especially crying, when excessive motion artefact was most likely.

Conclusions: The data suggest that Masimo signal extraction technology may offer improvement in pulse oximetry performance, particularly in clinical situations in which extreme motion artefacts are likely.

Keywords: oxygen saturation; heart rate; signal extraction technology

Abbreviations: HR, heart rate; ECG, electrocardiogram; SpO2, oxygen saturation


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Fantoms
Ben Stenson
Arch. Dis. Child. Fetal Neonatal Ed. 2003 88: F448. [Extract] [Full Text] [PDF]

This article has been cited by other articles:

  • Quine, D, Stenson, B J (2008). Does the monitoring method influence stability of oxygenation in preterm infants? A randomised crossover study of saturation versus transcutaneous monitoring. Arch. Dis. Child. Fetal Neonatal Ed. 93: F347-F350 [Abstract] [Full Text]  
  • Geisheker, J V, Hill, G (2004). US bioethics fall short of world standards. Arch. Dis. Child. Fetal Neonatal Ed. 89: F563-F563 [Full Text]  

eLetters:

Read all eLetters

Exposure of neonates to unnecessary pain is unethical
Gregory J. Boyle, PhD (Melb & Delaware)
Fetal Neonatal Ed. Online, 12 Dec 2003 [Full text]

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Latest from ADC

 

ADC is co-owned by the RCPCH and is the official journal of the European Academy of Paediatrics

BMJ Careers - Latest Paediatrics and Paediatric Surgery Jobs

Paediatrics and Paediatric Surgery Jobs