Archives of Disease in Childhood - Fetal and Neonatal Edition 2004;89:F356-F359; doi:10.1136/adc.2003.033548
Copyright © 2004 BMJ Publishing Group Ltd & Royal College of Paediatrics and Child Health.
Archives of Disease in Childhood Fetal and Neonatal Edition 2004;89:F356
© 2004 Archives of Disease in Childhood Fetal and Neonatal Edition

ORIGINAL ARTICLE

Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice

H T Ho, W K Y Yeung and B W Y Young

Department of Paediatrics and Adolescent Medicine, Pamela Youde Nethersole Eastern Hospital, Hong Kong SAR, China

Correspondence to:
Correspondence to:
Dr Ho
Department of Paediatrics and Adolescent Medicine, Pamela Youde Nethersole Eastern Hospital, 3 Lok Man Road, Chai Wan, Hong Kong SAR, China; brianyon{at}netvigator.com

Background: Low blood glucose in newborns is difficult to detect clinically. Hence a reliable "point of care" device (glucometer) for early detection and treatment of low glucose is needed.

Objective: To evaluate the performance of five readily available glucometers for the detection of low blood glucose in newborn infants.

Method: Glucostix measurements were taken for newborns with risk factors using a Reflolux S (Boehringer) glucometer. If the initial reading was low (< 2.6 mmol/l), further measurements were taken with two other glucometers (phase I, Advantage and Glucotrend (Roche); phase II, Elite XL (Bayer) and Precision (Abbott)), and plasma glucose was measured in the laboratory (Aeroset; Abbott).

Results: Over 10 months, 101 specimens were collected from 71 newborns (57 in phase I; 44 in phase II). The Advantage glucometer usually overestimated blood glucose with a mean difference of 1.07 mmol/l (p < 0.01) at all low glucose ranges. The Glucotrend, Precision, and Elite XL glucometers performed better; the mean differences were not significantly different from the laboratory measured value (0.17 mmol/l (p  =  0.37); –0.12 mmol/l (p  =  0.13), and 0.24 mmol/l (p  =  0.13) respectively). For detection of glucose concentrations < 2.6 mmol/l, the Precision glucometer had the highest sensitivity (96.4%) and negative predictive value (90%). For lower glucose concentrations (< 2.0 mmol/l), the Glucotrend glucometer performed even better (sensitivity 92.3%, negative predictive value 96.3%).

Conclusion: Point of care devices should have good precision in the low glucose concentration range, sensitivity, and accuracy for early detection of neonatal hypoglycaemia. None of the five glucometers was satisfactory as the sole measuring device. The Glucotrend and Precision glucometers have the greatest sensitivity and negative predictive value. However, confirmation with laboratory measurements of plasma glucose and clinical assessment are still of the utmost importance.

Abbreviations: ADA, American Diabetic Association; NCCLS, National Committee for Clinical Laboratory Standards; NPV, negative predictive value; PPV, positive predictive value

Keywords: glucose; glucometer; hypoglycaemia; point of care device


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

This article has been cited by other articles:

  • Mahoney, J., Ellison, J. (2007). Assessing the Quality of Glucose Monitor Studies: A Critical Evaluation of Published Reports. Clin. Chem. 53: 1122-1128 [Abstract] [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Latest from ADC

 

ADC is co-owned by the RCPCH and is the official journal of the European Academy of Paediatrics

BMJ Careers - Latest Paediatrics and Paediatric Surgery Jobs

Paediatrics and Paediatric Surgery Jobs