Archives of Disease in Childhood - Fetal and Neonatal Edition 2006;91:F193-F196
ORIGINAL ARTICLE
Plethysmograph and interrupter resistance measurements in prematurely born young children
1 Department of Child Health, Guys, Kings and St Thomas Medical School, Kings College London, UK
2 Department of Child Health, St Georges Hospital Medical School, London, UK
3 School of Health Sciences and Social Care, Brunel University, West London, UK
4 Academic Division of Child Health, School of Human Development, University of Nottingham, Nottingham, UK
Correspondence to:
Correspondence to:
Professor Greenough
Department of Child Health, 4th Floor Golden Jubilee Wing, Kings College Hospital, Bessemer Road, London SE5 9RS, UK; anne.greenough{at}kcl.ac.uk
Background: Airways obstruction in premature infants is often assessed by plethysmography, which requires sedation. The interrupter (Rint) technique does not require sedation, but has rarely been examined in children under 2 years of age.
Objective: To compare Rint results with plethysmographic measurements of airway resistance (Raw) in prematurely born, young children.
Design: Prospective study.
Setting: Infant and Paediatric Lung Function Laboratories.
Patients: Thirty children with a median gestational age of 2529 weeks and median postnatal age of 13 months.
Interventions and main outcome measures: The infants were sedated, airway resistance was measured by total body plethysmography (Raw), and Rint measurements were made using a MicroRint device. Further Raw and Rint measurements were made after salbutamol administration if the children remained asleep.
Results: Baseline measurements of Raw and Rint were obtained from 30 and 26 respectively of the children. Mean baseline Rint values were higher than mean baseline Raw results (3.45 v 2.84 kPa/l/s, p = 0.006). Limits of agreement for the mean difference between Rint and Raw were 1.52 to 2.74 kPa/l/s. Ten infants received salbutamol, after which the mean Rint result was 3.6 kPa/l/s and mean Raw was 3.1 kPa/l/s (limits of agreement 0.28 to 1.44 kPa/l/s).
Conclusion: The poor agreement between Rint and Raw results suggests that Rint measurements cannot substitute for plethysmographic measurements in sedated prematurely born infants.
Abbreviations: Raw, airways resistance measured by total body plethysmography; Rint, interrupter resistance
Keywords: airways resistance; interrupter technique; premature; plethysmography
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.