Arch. Dis. Child

HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER
[Advanced]

Published Online First: 18 December 2007. doi:10.1136/adc.2007.129098
Archives of Disease in Childhood - Fetal and Neonatal Edition 2008;93:F132-F134
Copyright © 2008 BMJ Publishing Group Ltd & Royal College of Paediatrics and Child Health

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
adc.2007.129098v1
93/2/F132    most recent
Right arrow Submit a response
Right arrow Read responses to this article
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this link to a friend
Right arrow Similar articles in ADC Online
Right arrow Similar articles in PubMed
Right arrow Add article to my folders
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Karam, O
Right arrow Articles by Rimensberger, P C
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Karam, O
Right arrow Articles by Rimensberger, P C
Topic Collections
Right arrowRelevant Article

ORIGINAL ARTICLES

Noise levels during nCPAP are flow-dependent but not device-dependent

O Karam1, C Donatiello1, E Van Lancker2, V Chritin2, R E Pfister1, P C Rimensberger1

1 Pediatric and Neonatology Critical Care Unit, Geneva’s University Hospital, Geneva, Switzerland
2 IAV Engineering, Ecole Polytechnique Fédérale de Lausanne, Switzerland

Correspondence to:
Dr Oliver Karam, Soins Intensifs de Pédiatrie, Hôpital des Enfants, 6, rue Willy Donzé, 1211 Genève, Switzerland; karam{at}drugdoses.net

Objective: Nasal continuous positive airway pressure (nCPAP) has been shown to improve the outcome of infants with respiratory distress syndrome. However, noise generation could be of concern. Therefore, our study was designed to measure the noise levels of various CPAP drivers.

Design: For infants admitted to our neonatal intensive care unit and paediatric critical care unit, we measured the noise level in the oral cavity, using a microphonic probe with a flexible capillary tube. Various CPAP drivers and interfaces have been tested.

Results: 27 measurements were made in eight infants. Mean noise level was 88.6 (SD 18.8) dB and was correlated with flow (p<0.01) but not with pressure. A noise level above 90 dB was detected in 67% of the measurements.

Conclusions: nCPAP drivers are valuable devices for neonatal care that may prevent primary mechanical ventilation or re-intubation, but generate a large amount of noise, often higher than occupational limits accepted for adult workers. Therefore, new devices must be designed to minimise this possible noxious exposure of premature infants to unacceptably high noise levels.



Relevant Article

Fantoms
Martin Ward Platt, Deputy Edi
Arch. Dis. Child. Fetal Neonatal Ed. 2008 93: F79. [Extract] [Full Text] [PDF]



This article has been cited by other articles:


Home page
Arch. Dis. Child. Fetal Neonatal Ed.Home page
D Trevisanuto, L Camiletti, A Udilano, N Doglioni, and V Zanardo
Noise levels during neonatal helmet CPAP
Arch. Dis. Child. Fetal Neonatal Ed., September 1, 2008; 93(5): F396 - F397.
[Full Text] [PDF]

eLetters:

Read all eLetters

Noise levels during neonatal helmet CPAP
Daniele Trevisanuto, et al.
Fetal Neonatal Ed. Online, 25 Mar 2008 [Full text]



HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER
ARCH DIS CHILD FETAL NEONATAL ED ED PRACTICE
Terms and conditions relating to subscriptions purchased online  ¦  Website terms and conditions  ¦  Privacy policy
Copyright © 2008 BMJ Publishing Group Ltd & Royal College of Paediatrics and Child Health