Noise levels during nCPAP are flow-dependent but not device-dependent

O Karam1, C Donatiello1, E Van Lancker2, V Chritin2, R E Pfister1 and P C Rimensberger1

1 Pediatric and Neonatology Critical Care Unit, Geneva’s University Hospital, Geneva, Switzerland
2 IAV Engineering, Ecole Polytechnique Fédérale de Lausanne, Switzerland

Correspondence to:
Dr Oliver Karam, Soins Intensifs de Pédiatrie, Hôpital des Enfants, 6, rue Willy Donzé, 1211 Genève, Switzerland; karam{at}drugdoses.net

Objective: Nasal continuous positive airway pressure (nCPAP) has been shown to improve the outcome of infants with respiratory distress syndrome. However, noise generation could be of concern. Therefore, our study was designed to measure the noise levels of various CPAP drivers.

Design: For infants admitted to our neonatal intensive care unit and paediatric critical care unit, we measured the noise level in the oral cavity, using a microphonic probe with a flexible capillary tube. Various CPAP drivers and interfaces have been tested.

Results: 27 measurements were made in eight infants. Mean noise level was 88.6 (SD 18.8) dB and was correlated with flow (p<0.01) but not with pressure. A noise level above 90 dB was detected in 67% of the measurements.

Conclusions: nCPAP drivers are valuable devices for neonatal care that may prevent primary mechanical ventilation or re-intubation, but generate a large amount of noise, often higher than occupational limits accepted for adult workers. Therefore, new devices must be designed to minimise this possible noxious exposure of premature infants to unacceptably high noise levels.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Fantoms
Martin Ward Platt, Deputy Edi
Arch. Dis. Child. Fetal Neonatal Ed. 2008 93: F79. [Extract] [Full Text] [PDF]

This article has been cited by other articles:

  • Trevisanuto, D, Camiletti, L, Udilano, A, Doglioni, N, Zanardo, V (2008). Noise levels during neonatal helmet CPAP. Arch. Dis. Child. Fetal Neonatal Ed. 93: F396-F397 [Full Text]  

eLetters:

Read all eLetters

Noise levels during neonatal helmet CPAP
Daniele Trevisanuto, et al.
Fetal Neonatal Ed. Online, 25 Mar 2008 [Full text]

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Latest from ADC

 

ADC is co-owned by the RCPCH and is the official journal of the European Academy of Paediatrics

BMJ Careers - Latest Paediatrics and Paediatric Surgery Jobs

Paediatrics and Paediatric Surgery Jobs