Arch Dis Child Fetal Neonatal Ed 78:F148-F150 doi:10.1136/fn.78.2.F148
  • Original article

Volume delivery during high frequency oscillation

Table 1

Comparison of the infants’ demographics, oscillatory settings and delivered volume according to oscillator type, data are median (range)

SensorMedics (n=28) SLE (n=17)
Gestational age (weeks) 29 (24–33) 31 (27–37)
MAP (cmH2O) 12 (10–25) 11 (8–17)
Oscillatory amplitude (cmH2O) 24 (16–39) 33 (20–65)
OI 9.5 (3–21) 7 (4–15)
PaCO2 (mm Hg) 39 (27–50) 45 (27–37)
PaO2 (mm Hg) 61 (50–99) 63 (53–89)
Delivered volume (ml/kg) 2.2 (1.3–3.1) 2.6 (1.0–3.6)