Arch Dis Child Fetal Neonatal Ed 2001;84:F53-F55 doi:10.1136/fn.84.1.F53
  • Original article

An automatic incision device for obtaining blood samples from the heels of preterm infants causes less damage than a conventional manual lancet

  1. H Vertanena,
  2. V Fellmanb,
  3. M Brommelsc,
  4. L Viinikkaa
  1. aDepartment of Clinical Chemistry, Helsinki University Central Hospital, Helsinki, Finland, bHospital for Children and Adolescents, Helsinki University Central Hospital, cDepartment of Public Health, University of Helsinki, Helsinki
  1. Ms Vertanen, Department of Clinical Chemistry, Hospital for Children and Adolescents, PO Box 281, FIN-00029 HUS, Finland email: helena.vertanen{at}hus.fi
  • Accepted 8 April 2000

Abstract

OBJECTIVES To evaluate in a randomised blind study the effect on puncture site lesions of two different incision devices used to obtain blood samples from preterm infants by repeated heel sticks.

SETTING The neonatal intensive care unit at the Hospital for Children and Adolescents and Laboratory, Helsinki University Central Hospital.

PATIENTS A total of 100 preterm infants (birth weight below 2500 g) not previously subjected to heel stick sampling.

INTERVENTIONS The infants were randomly allocated to blood sampling from the heel with either a conventional manual lancet or an automatic incision device. The same type of lancet was used for any given baby throughout the study (2–21 days).

MAIN OUTCOME MEASURES The damage caused by sampling was evaluated using four criteria: bruising of the heel, inflammation of the heel, bruising of either the ankle or the leg, and skin healing at the puncture site. The evaluation was based on photographs presenting typical categories of each outcome.

RESULTS To obtain a sufficient volume of blood, on average 2.6 times more punctures were needed when the conventional manual lancet was used than when the automatic incision device was used. Heels punctured with the lancet had more bruising (100% v 84%) and more signs of inflammation (79% v 53%), and there was more bruising of the ankle or leg (92% v53%) than when the automatic incision device was used. Skin healed equally rapidly in the two groups.

CONCLUSION The use of an automatic incision device for collecting repeated skin puncture samples from preterm infants is less traumatic than the use of a conventional manual lancet.

Footnotes

    Register for free content

    The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.