Arch Dis Child Fetal Neonatal Ed 87:F137-F140 doi:10.1136/fn.87.2.F137
  • Original article

A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields

Table 3

Specificity and positive predictive values for respective devices when sensitivity is set to 100%

SBR ≥250 μmol/l SBR <250 μmol/l Totals Positive predictive value
SBR, Serum bilirubin; TcB, transcutaneous bilirubinometry.
Minolta
    TcB reading ≥170 μmol/l 17 32 49 34.7
    TcB reading <170 μmol/l 0 15 15
    Totals 17 47 64
100.0 (sensitivity) 31.9 (specificity)
SpectRx A
    TcB reading ≥150 μmol/l 17 37 54 31.5
    TcB reading <150 μmol/l 0 10 10
    Totals 17 47 64
100.0 (sensitivity) 21.3 (specificity)
SpectRx B
    TcB reading ≥150 μmol/l 17 34 51 33.3
    TcB reading <150 μmol/l 0 13 13
    Totals 17 47 64
100.0 (sensitivity) 27.7 (specificity)