rss
Arch Dis Child Fetal Neonatal Ed 88:F315-F318 doi:10.1136/fn.88.4.F315
  • Original article

Noise levels within the ear and post-nasal space in neonates in intensive care

Table 4

Mean noise levels in the post-nasal space for different flow rates of CPAP

Flow rate (litres/min) 0.2 kHz 0.5 kHz 1 kHz 2 kHz 4 kHz 6 kHz 8 kHz
The number of individual measurements for each flow rate are shown in parentheses.
CPAP, Continuous positive airways pressure.
5 (6) 86.57 86.43 88.75 76.43 80.48 79.68 72.42
6 (7) 86.74 89.30 87.90 77.2 75.09 72.83 66.41
7 (7) 91.93 91.57 91.27 81.79 73.01 75.14 67.47
8 (7) 90.64 93.18 93.72 82.61 80.89 78.50 73.37
9 (7) 94.59 96.93 96.87 87.44 82.61 78.38 76.6
10 (7) 101.9 102.2 100.3 90.84 90.04 88.51 84.83