rss
Arch Dis Child Fetal Neonatal Ed 88:F505-F508 doi:10.1136/fn.88.6.F505
  • Original article

Motion resistant pulse oximetry in neonates

Table 3

Relation of Masimo and Nellcor oximeter heart rates to cardiac monitor heart rate before, during, and after circumcision (n=15)

Baseline Circumcision After circumcision
Masimo Nellcor Masimo Nellcor Masimo Nellcor
Correlation coefficient (r2) 0.86 0.63 0.72 0.28 0.78 0.31
Residual error (beats/min) 1.3 3.9 5.8 19.5 2.4 14.6

This Article