Arch Dis Child Fetal Neonatal Ed 2004;89:F174-F176 doi:10.1136/adc.2002.023937
  • Original article

Accuracy of the volume and pressure displays of high frequency oscillators

  1. J A Leipälä,
  2. S Iwasaki,
  3. A Milner,
  4. A Greenough
  1. Children Nationwide Neonatal Intensive Care Unit, King’s College Hospital, London, UK
  1. Correspondence to:
    Professor Greenough
    Department of Child Health, King’s College Hospital, London SE5 9PJ, UK; anne.greenoughkcl.ac.uk
  • Accepted 11 March 2003

Abstract

Objective: To determine the effect of frequency on the accuracy of volume and pressure displays of high frequency oscillators.

Methods: The effect of frequency on the displayed volume of the Stephanie, Dräger Babylog 8000 Plus, and SLE 5000 oscillators was assessed. A sine wave pump delivered a constant tidal volume at frequencies of 5–15 Hz to the patient manifold of the oscillators. The displayed volumes at each frequency were compared with the delivered volume. The effect of frequency on displayed pressure was assessed by connecting the oscillator’s patient manifold to a lung model; three types of oscillator were studied (SensorMedics 3100A, SLE 5000, and Stephanie). Airway pressure was measured from the manifold using a pressure transducer and non-compliant tubing; the pressure measuring system had a flat frequency response to 30 Hz.

Results: The SLE 5000 volume display overread the delivered volume (by about 5%), but was not affected by frequency. At 5 Hz, the Dräger Babylog 8000 Plus and the Stephanie underread the delivered volume (by about 20%). Increasing frequency resulted in a greater discrepancy between the delivered and displayed volume with the Stephanie, but a smaller discrepancy with the Dräger Babylog 8000 Plus. Altering frequency had a small effect (maximum difference 6%) on the relation between the displayed and delivered pressure for all three oscillators.

Conclusion: Frequency affects the accuracy of displayed volumes and, to a lesser extent, displayed pressures of high frequency oscillators. The results emphasise that data displayed by new devices should not be uncritically accepted.

Footnotes

    This Article

    Services

    1. Request permissions

    Responses

    1. Submit a response
    2. No responses published

    Social bookmarking

    Latest from Education & Practice

    Latest from Education & Practice

    Register for free content

    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of ADC Fetal & Neonatal.
    View free sample issue >>

    Free archive
    The full back archive is now available for ADC Fetal & Neonatal. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

  • Paediatrics and Paediatric Surgery Jobs

    Paediatrics and Paediatric Surgery Jobs