Arch Dis Child Fetal Neonatal Ed 2004;89:F356-F359 doi:10.1136/adc.2003.033548
  • Original article

Evaluation of “point of care” devices in the measurement of low blood glucose in neonatal practice

  1. H T Ho,
  2. W K Y Yeung,
  3. B W Y Young
  1. Department of Paediatrics and Adolescent Medicine, Pamela Youde Nethersole Eastern Hospital, Hong Kong SAR, China
  1. Correspondence to:
    Dr Ho
    Department of Paediatrics and Adolescent Medicine, Pamela Youde Nethersole Eastern Hospital, 3 Lok Man Road, Chai Wan, Hong Kong SAR, China; brianyonnetvigator.com
  • Accepted 15 September 2003

Abstract

Background: Low blood glucose in newborns is difficult to detect clinically. Hence a reliable “point of care” device (glucometer) for early detection and treatment of low glucose is needed.

Objective: To evaluate the performance of five readily available glucometers for the detection of low blood glucose in newborn infants.

Method: Glucostix measurements were taken for newborns with risk factors using a Reflolux S (Boehringer) glucometer. If the initial reading was low (< 2.6 mmol/l), further measurements were taken with two other glucometers (phase I, Advantage and Glucotrend (Roche); phase II, Elite XL (Bayer) and Precision (Abbott)), and plasma glucose was measured in the laboratory (Aeroset; Abbott).

Results: Over 10 months, 101 specimens were collected from 71 newborns (57 in phase I; 44 in phase II). The Advantage glucometer usually overestimated blood glucose with a mean difference of 1.07 mmol/l (p < 0.01) at all low glucose ranges. The Glucotrend, Precision, and Elite XL glucometers performed better; the mean differences were not significantly different from the laboratory measured value (0.17 mmol/l (p  =  0.37); −0.12 mmol/l (p  =  0.13), and 0.24 mmol/l (p  =  0.13) respectively). For detection of glucose concentrations < 2.6 mmol/l, the Precision glucometer had the highest sensitivity (96.4%) and negative predictive value (90%). For lower glucose concentrations (< 2.0 mmol/l), the Glucotrend glucometer performed even better (sensitivity 92.3%, negative predictive value 96.3%).

Conclusion: Point of care devices should have good precision in the low glucose concentration range, sensitivity, and accuracy for early detection of neonatal hypoglycaemia. None of the five glucometers was satisfactory as the sole measuring device. The Glucotrend and Precision glucometers have the greatest sensitivity and negative predictive value. However, confirmation with laboratory measurements of plasma glucose and clinical assessment are still of the utmost importance.

Footnotes

    This Article

    Services

    1. Request permissions

    Responses

    1. Submit a response
    2. No responses published

    Social bookmarking

    Latest from Education & Practice

    Latest from Education & Practice

    Register for free content

    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of ADC Fetal & Neonatal.
    View free sample issue >>

    Free archive
    The full back archive is now available for ADC Fetal & Neonatal. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

  • Paediatrics and Paediatric Surgery Jobs

    Paediatrics and Paediatric Surgery Jobs