Arch Dis Child Fetal Neonatal Ed 2004;89:F490-F493 doi:10.1136/adc.2003.047712
  • Neonatal resuscitation

Comparison of three manual ventilation devices using an intubated mannequin

  1. S G Hussey,
  2. C A Ryan,
  3. B P Murphy
  1. Erinville Hospital, Cork, Ireland
  1. Correspondence to:
    Professor Ryan
    Department of Paediatrics and Child Health, Erinville Hospital, Cork, Ireland; ryant1shb.ie
  • Accepted 11 May 2004

Abstract

Objective: To compare three devices for manual neonatal ventilation.

Design: Participants performed a two minute period of ventilation using a self inflating device, an anaesthesia bag with attached manometer, and a Neopuff device. An intubated neonatal mannequin, approximating a 1 kg infant with functional lungs, was used for the study. Target ventilation variables included a rate of 40 breaths per minute, peak inspiratory pressure (PIP) of 20 cm H2O, and positive end expiratory pressure (PEEP) of 4 cm H2O. The circuit was attached to a laptop computer for data recording.

Results: Thirty five participants were enrolled, including consultant neonatologists, paediatricians, and anaesthetists, paediatric and anaesthetic registrars, and neonatal nurses. The maximum PIP recorded using the self inflating bag, anaesthetic bag, and Neopuff device were 75.9, 35.5, and 22.4 cm H2O respectively. There were significant differences between the devices for mean PIP (30.7, 18.1, and 20.1 cm H2O), mean PEEP (0.2, 2.8, and 4.4 cm H2O), mean airway pressure (7.6, 8.5, and 10.9 cm H2O), % total breaths ≤ 21 cm H2O PIP (39%, 92%, and 98%), and % total breaths ≥ 30 cm H2O PIP (45%, 0, and 0). There was no difference between doctors and allied health professionals for the variables examined.

Conclusion: The anaesthetic bag with manometer and Neopuff device both facilitate accurate and reproducible manual ventilation. Self inflating devices without modifications are not as consistent by comparison and should incorporate a manometer and a PEEP device, particularly when used for resuscitation of very low birthweight infants.

Footnotes

  • Fisher & Paykel supplied the Neopuff device, mannequin, and recording software for this study. The authors have neither financial interests nor other relationships with the companies whose products or services are discussed.

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.