rss
Arch Dis Child Fetal Neonatal Ed 89:F490-F493 doi:10.1136/adc.2003.047712
  • Neonatal resuscitation

Comparison of three manual ventilation devices using an intubated mannequin

Table 3

 Ventilatory variables

Response Self inflating bag Anaesthetic bag Neopuff p Value
Values are mean (SEM). Values followed by an asterisk within the same row are not significantly different (p > 0.05).
PEEP, Positive end expiratory pressure; PIP, peak inspiratory pressure.
Mean max PIP 44.7 (2.3) 22.6 (0.7) 20.4 (0.5) <0.001
Mean PIP 30.7 (1.9) 18.1 (0.4) 20.1 (0.1) <0.001
Mean PEEP 0.15 (0.03) 2.83 (0.23) 4.41 (0.08) <0.001
Mean airway pressure 7.6 (0.8)* 8.5 (0.3)* 10.9 (0.3) <0.001
Mean rate 47.1 (3.0)* 47.3 (2.7)* 39.7 (1.8) <0.05
% total breaths ⩽21 cm H2O PIP 39 (0.07) 92 (0.02)* 98 (0.02)* <0.001
% total breaths ⩾30 cm H2O PIP 45 (0.07) 0 * 0* <0.001