Arch Dis Child Fetal Neonatal Ed 2005;90:F397-F400 doi:10.1136/adc.2004.064709
  • Original article

Neonatal resuscitation 3: manometer use in a model of face mask ventilation

  1. C P F O’Donnell1,
  2. P G Davis1,
  3. R Lau1,
  4. P A Dargaville2,
  5. L W Doyle1,
  6. C J Morley1
  1. 1Royal Women’s Hospital Melbourne, Victoria 3053, Australia
  2. 2Murdoch Children’s Research Institute, Melbourne, Victoria 3052, Australia
  1. Correspondence to:
    Dr O’Donnell
    Royal Women’s Hospital Melbourne, 132 Grattan Street, Carlton, Victoria 3053, Australia; colm.odonnellrwh.org.au
  • Accepted 18 March 2005
  • Published Online First 4 May 2005

Abstract

Background: Adequate ventilation is the key to successful neonatal resuscitation. Positive pressure ventilation (PPV) is initiated with manual ventilation devices via face masks. These devices may be used with a manometer to measure airway pressures delivered. The expiratory tidal volume measured at the mask (VTE(mask)) is a good estimate of the tidal volume delivered during simulated neonatal resuscitation.

Aim: To assess the effect of viewing a manometer on the peak inspiratory pressures used, the volume delivered, and leakage from the face mask during PPV with two manual ventilation devices in a model of neonatal resuscitation.

Methods: Participants gave PPV to a modified resuscitation mannequin using a Laerdal infant resuscitator and a Neopuff infant resuscitator at specified pressures ensuring adequate chest wall excursion. Each participant gave PPV to the mannequin with each device twice, viewing the manometer on one occasion and unable to see the manometer on the other. Data from participants were averaged for each device used with the manometer and without the manometer separately.

Results: A total of 7767 inflations delivered by the 18 participants were recorded and analysed. Peak inspiratory pressures delivered were lower with the Laerdal device. There were no differences in leakage from the face mask or volumes delivered. Whether or not the manometer was visible made no difference to any measured variable.

Conclusions: Viewing a manometer during PPV in this model of neonatal resuscitation does not affect the airway pressure or tidal volumes delivered or the degree of leakage from the face mask.

Footnotes

  • Published Online First 4 May 2005

  • Competing interests: none declared

Latest from Education & Practice

Latest from Education & Practice

Register for free content

Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of ADC Fetal & Neonatal.
View free sample issue >>

Free archive
The full back archive is now available for ADC Fetal & Neonatal. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

  • Paediatrics and Paediatric Surgery Jobs

    Paediatrics and Paediatric Surgery Jobs