Arch Dis Child Fetal Neonatal Ed 93:F302-F304 doi:10.1136/adc.2007.120428
  • Short report

Pharyngeal pressure value using two continuous positive airway pressure devices

  1. M Colnaghi,
  2. P G Matassa,
  3. M Fumagalli,
  4. D Messina,
  5. F Mosca
  1. 1
    NICU-Fondazione IRCCS, Ospedale Maggiore Policlinico, Mangiagalli e Regina Elena, Universitè degli Studi di Milano, Milan, Italy
  1. Professor F Mosca, NICU-Fondazione IRCCS, Ospedale Maggiore Policlinico, Mangiagalli e Regina Elena, Via della Commenda 12, 20122 Milan, Italy; fabio.mosca{at}unimi.it
  • Accepted 14 November 2007
  • Published Online First 11 March 2008

Abstract

The aim of the study was to measure the difference between the set continuous positive airway pressure (CPAP) value and the pharyngeal pressure reading during CPAP in premature infants with mild respiratory distress syndrome, using two different devices: hood CPAP and the conventional nasal system. The preliminary results suggest that hood CPAP may produce more stable pharyngeal pressure than the conventional nasal device.

Footnotes

  • Competing interests: None.

  • Ethics approval: Parental/guardian informed consent was obtained for publication of fig 1.

  • Patient consent: Ethics approval was obtained.

Latest from Education & Practice

Latest from Education & Practice

Register for free content

Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of ADC Fetal & Neonatal.
View free sample issue >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Paediatrics and Paediatric Surgery Jobs

Paediatrics and Paediatric Surgery Jobs