Arch Dis Child Fetal Neonatal Ed 2008;93:F302-F304 doi:10.1136/adc.2007.120428
  • Short report

Pharyngeal pressure value using two continuous positive airway pressure devices

  1. M Colnaghi,
  2. P G Matassa,
  3. M Fumagalli,
  4. D Messina,
  5. F Mosca
  1. 1
    NICU-Fondazione IRCCS, Ospedale Maggiore Policlinico, Mangiagalli e Regina Elena, Universitè degli Studi di Milano, Milan, Italy
  1. Professor F Mosca, NICU-Fondazione IRCCS, Ospedale Maggiore Policlinico, Mangiagalli e Regina Elena, Via della Commenda 12, 20122 Milan, Italy; fabio.mosca{at}unimi.it
  • Accepted 14 November 2007
  • Published Online First 11 March 2008

Abstract

The aim of the study was to measure the difference between the set continuous positive airway pressure (CPAP) value and the pharyngeal pressure reading during CPAP in premature infants with mild respiratory distress syndrome, using two different devices: hood CPAP and the conventional nasal system. The preliminary results suggest that hood CPAP may produce more stable pharyngeal pressure than the conventional nasal device.

Footnotes

  • Competing interests: None.

  • Ethics approval: Parental/guardian informed consent was obtained for publication of fig 1.

  • Patient consent: Ethics approval was obtained.

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.