Potential hazard of the Neopuff T-piece resuscitator in the absence of flow limitation.

  1. Colin P Hawkes (cphawkes{at}gmail.com)
  1. Department of Neonatology, Cork University Maternity Hospital, Republic of Ireland
    1. Oluwole A Oni (wolexza{at}yahoo.com)
    1. Department of Neonatology, Cork University Maternity Hospital, Republic of Ireland
      1. Eugene M Dempsey (gene.dempsey{at}mailp.hse.ie)
      1. Department of Neonatology, Cork University Maternity Hospital, Republic of Ireland
        1. C Anthony Ryan (tonyryan007{at}gmail.com)
        1. Department of Neonatology, Cork University Maternity Hospital, Republic of Ireland
          • Published Online First 8 April 2009

          Abstract

          Objective: To assess (i) Peak Inspiratory Pressure (PIP), Positive End Expiratory Pressure (PEEP) and maximum pressure relief (Pmax) at different rates of gas flow, when the Neopuff had been set to function at 5L/min. (ii) To assess maximum PIP and PEEP at a flow rate of 10L/min with a simulated air leak of 50%.

          Design: 5 Neopuffs were set to a PIP of 20, PEEP of 5 and Pmax of 30 cm H2O at a gas flow of 5L/min. PIP, PEEP and Pmax were recorded at flow rates of 10, 15L/min and maximum flow. Maximum achievable pressures at 10L/min gas flow, with a 50% air leak, were measured.

          Results: At gas flow of 15L/min, mean PEEP increased to 20 (95% CI 20-21), PIP to 28 (95% CI 28-29) and the Pmax to 40 cm H2O (95% CI 38-42). At maximum flow (85L/min) a PEEP of 79 (95% CI 51-91) and PIP of 103 cm H2O (95% CI 69-115) were generated. At 10L/min flow, with an air leak of 50%, the maximum PEEP and PIP were 21 (95% CI 19-23) and 69 cm H2O (95% CI 66-71).

          Conclusions: The maximum pressure relief valve is overidden by increasing the rate of gas flow and potentially harmful PIP and PEEP can be generated. Even in the presence of a 50% gas leak, more than adequate pressures can be provided at 10L/min gas flow. We recommend the limitation of gas flow to a rate of 10L/min as an added safety mechanism for this device.

          Latest from Education & Practice

          Latest from Education & Practice

          Register for free content

          Free sample
          This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of ADC Fetal & Neonatal.
          View free sample issue >>

          Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

          Paediatrics and Paediatric Surgery Jobs

          Paediatrics and Paediatric Surgery Jobs